Error Compensation Current Transformer
(Standard Current Transformer)
TYPE 6460A
Model 6460A can output secondary current against primary current(5A) ratio error 10ppm and phase angle error 10µrad in accuracy where setting range is within 0-120%. It is convenient to expand the area of current value in every calibration works. Also, using it with Model 6462A as a set, you can create more expanded current range.
1.Characteristics
1.1 High accurate secondary current can be gained against setting value due to two-step structured transformer with compensating coil.
1.2 Secondary current can be output in one percent resolution accuracy in the area 0-120%.
1.3 Remotely controllable via standard IEEE-488 interface.
1.4 Model 6460A is durable for long term use because it is contained in aluminum-alloyed system case and transformer is molded.
2.Specifications
2.1 Primary Current 5A(Maximum10A)
2.2 Secondary Current Within 0 - 120% against primary current(Resolution 1%)
2.3 Frequency 50Hz to 400Hz
2.4 Ratio Error 10ppm
2.5 Phase Angle Error 10µrad
2.6 Rated Burden 25VA
2.7 GPIB Interface Equipped as standard
2.8 Power Supply 100V ac (50Hz to 60Hz)
2.9 Dimensions(W/H/D) 450 / 130 / 430 (mm)
2.10 Weight About15Kg
  • Model 6460A was developed under the technical guidance of JEMIC(Japan Electric Meters Inspection Corporation). Production and sales of this model is performed in accordance with a contract with them.
  • SunJEM accepts to produce other products with customer's specification.
  • Specification are subject to change without notice.